Turb® 750 T Benchtop Turbidimeter

Turb 750 T Overview | Benchtop Turbidimeter

Turb 750 T Operation | Benchtop Turbidimeter

Price$3,952.00 (US Only)

Option

SKU600131Y


In Stock 0 - Available for Backorder

Simplify turbidity measurements without compromising data quality!

YSI’s Turb® 750 T benchtop turbidity meter provides accurate and precise measurements up to 1100 FNU/NTU at an affordable price. The turbidity meter complies with EPA 180.1 and includes analytical quality assurance (AQA), which allows for adjustable calibration intervals, calibration record of turbidity standards used, and GLP-compliant documentation via PC and LIMS.

  • Overview

    With advanced features like intelligent reproducibility and plausibility check (IRPC), this powerful benchtop turbidity instrument can quickly take multiple measurements, automatically evaluate the data and remove outliers, offering your laboratory optimal efficiency.

    These features and the affordability make the Turb® 750 T a great turbidity meter for environmental, chemical, petroleum, food, and beverage laboratories!

    exo multiparameter sonde data datasonde


     One of the most affordable turbidity meters without compromising quality.

     


      Complies with EPA 180.1 and GLP

     

    exo anti-fouling antifouling


      Superior accuracy and precision with Xylem’s IRPC feature.

     


      Compact turbidity meter, occupying minimal benchtop space in your lab.
      

     

    Sample-ID and Data Management with the YSI Turb 750 T

    Get advanced data storage with the Turb Data Software, free to download from our website. Give samples unique IDs (up to 12 digits, including special characters) for streamlined output from the meter to your PC. Importing data into Excel or your LIMS system has never been easier! Press the store button twice, and your CSV files are automatically exported.

    Standards for Clear Results!

    YSI’s AMCO Clear® standards have been optimized for the optics of the Turb 750 T. A successful calibration with AMCO Clear standards is essential to obtain precise and reproducible results. Utilization of commercial formazine standards can influence results – learn more about our standards.

  • Specifications
    Turb750 Specifications  
    Measuring standards US EPA 180.1
    Light source

    White light Tungsten filament lamp

    Measuring mode Nephelometric (90˚ scattered light)
    Display

    Backlit graphics display, 160 x 104 Pixels

    Keypad Easy-to-clean foil keypad with alphanumeric entry option

    Measuring range

    0 - 1100 NTU

    Units

    NTU

    Resolution

    0.01 FNU/NTU in the range 0.00 – 9.99 FNU/NTU

    0.1 FNU/NTU in the range 10 – 99.99 FNU/NTU

    1 FNU/NTU in the range 100 – 1100 FNU/NTU

    Accuracy

    0.01 NTU or ± 2% of reading,

    ± 3% in the range 500–1100 NTU

    Repeatability < 1% of reading
    Reading mode Measurement with Intelligent Reproducibility and Plausibility Check (IRPC) procedure, rapid settlement samples supported by fast response time and IRPC.
    Response time

    7 sec

    Calibration – options Default 3 points standard calibration, flexible calibration with 2–5 user-defined calibration points, QUICKCal
    Calibration protocol and interval setting

    yes/yes

    AQS-Support Calibration protocol, Analytical Quality Assurance (AQA) flag, cal flag
    Data storage

    2500 data sets with cal protocol, AQA flag

    Sample Identification Alphanumeric entry via keypad

    Firmware update

    via USB

    Interface

    RS232, USB, printer via PC or RS232

    Storage condition instrument

    – 25 to + 65 °C (13 to 149 °F)

    Operating temperature range

    +5 to + 55 °C (41 to 131 °F)

     +5 to + 40 °C (41…104 °F) with a power plug connected

    GLP-compliant PC software Turb® Data
    Dimensions (H x W x D) 290 x 190 x 80 mm (11.42 x 7.48 x 3.15 inches)
    Weight

    1.1 kg (0.22 lbs)

    Calibration standards

    Cal.Kit for 3-P standard calibration: long-term stable polymer AMCO® Clear standards (0.02, 10.0, 1000 FNU/NTU)

    Vials, sample volume

    28 mm diameter vials, min. volume 15 ml, borosilicate glass, phenolic resin cap, PTFE-coated rubber seal. No silicon oil is required to cover scratches for the measurement procedure!

    Sample conditions

    Sample temperature < 70 °C (158 °F)

    Power supply Wide range power supply with plugs for Euro, US, UK, and Australia
    Certificates CE
    Delivery scope

    Lab turbidimeter Turb® 750 T, four 1.5 V AA type batteries, wide-range power supply, cable USB-A to USB-B mini, six empty 28 mm vials, cap labels for orientation marking, Cal.Kit Turb® 750 T, cloth, quick guide, compact operation manual, extended manual, Turb® Data PC software, inspection protocol

  • Software
    Turb® Data – PC-Software for data management of the turbidimeter series Turb® 750 – Version 5.01 (English)

    Software to install on your PC for communication with Turb750 T instruments, allowing for data transfer.

    GLP-compliant data transfer in CSV format Instrument recognition and user name, Data import into Excel, Data import into LIMS, Cable USB-Mini – USB-A in delivery of Turb® 750.

    Download- and installation instructions: Unzip and open the file and follow the instructions

    System requirements: Microsoft® Windows 7, Windows 8, Windows 10


    Turb 750 IR/T - Software Update - V2.66

    Attention: All stored measurement data will be deleted! Please save data before executing the Update.

    Software to install on your PC for communication with Turb750 T instruments, allowing for data transfer.

    Alphanumeric ID including special characters, 12 digits Auto-Print for connected printers, Auto-print into opened terminal and LIMS programs.

    Get the Software

  • Video

    Turb 750 T Overview | Benchtop Turbidimeter

    Turb 750 T Operation | Benchtop Turbidimeter

You may also need

MultiLab 4010-1W

Single channel benchtop multiparameter instrument

$1,384.24

TitroLine® 7000

Advanced Titrator

$6,937.00 - $7,427.00

TitroLine® 7800

Universal Titrator with IDS technology

$8,257.00 - $8,350.00

TITRONIC® 500

Piston Burette for Manual Titrations

$2,721.00 - $3,156.00

MultiLab IDS ProOBOD

Optical Self stirring BOD probe with temperature sensor, IDS

$1,163.76 - $1,352.00