5200
Multiparameter Monitoring and Control Instrument
Designed specifically for Aquaculture Systems, the YSI 5200 and AquaManager™ Software integrate process control, alarming, and data management into one product. Simple enough to monitor one tank, powerful enough to manage a full scale farming operation from anywhere in the world. Simultaneously measure dissolved oxygen, pH, conductivity, ORP, salinity and temperature.
YSI has Authorized Distributors all over the globe. View a map and find your closest contact for YSI 5200 products and accessories.
Have a question about your YSI 5200? Check out the support page with resources compiled just for you.
Want to return a product for service? Here you’ll find authorized service centers and instructions on how to send in your YSI 5200.
Optional AquaManager™ Windows™-based software links data management, system management, and networking to the desktop and internet. Browser-based format makes navigation, setup and maintenance easy.
The 5200 is a fully scaleable system - simply add units as your operation grows.
Features
Pond monitoring has never been easier with the 5200 and the 5567 wiper/stirrer. Save energy and maintenance with the 5200 and the 5567 controlling your paddle wheels or aeration system.
Recent Upgrades
YSI 5200 Instrument General Specifications
| Auxiliary Inputs |
2 ea inputs - 1 selectable digital or analog (0 to 1.0 VDC or 0 to 5.0 VDC selectable) 1 ea digital (contact open/close w/current source) | |
| Certifications | UL, CUL and CE | |
| Communication Ports | 1 ea RS-485 Full Duplex 1 ea RS-232-C Serial Communication | |
| Data Logger |
32K Memory – 750 sample data sets User selectable save rate | |
| Dimensions | 6.5 x 8.5 x 4.75 in (h x l x d) |
|
| Display |
2 x 16 character, extended temperature range LCD with backlight | |
| Instrument Operating Ranges |
Enclosure |
NEMA 4X Enclosure |
| Environment | Direct exposure to outside conditions | |
| Temperature | - 15 to 70ºC (23 to 158 ºF) | |
| Output Relays |
4; 5 amp 230 VAC (10 amp 120 VAC) relays for parameter, control and/or alarm | |
| Power Requirements |
Nominal | 12 VDC |
| Operating Range | 8.5 to 16.5 VDC (below 8.5 VDC uses optional battery back up) |
|
| Operating Current | 500 - 800 mA depending on mode of operation | |
| Software |
AquaManager | Windows 95, 98, ME, NT 3.51, NT 4.0, 2000, and XP |
| Autopoller | Windows 2000 and higher | |
| Timers |
4 ea General purpose timers, 1 start & 1 stop time per 24 hour period |
|
| Warranty | 2-year on instrument 1-year on probes and cable | |
YSI 5200 System Specifications (Instrument with Cable & Probe)
| Sensor Type | Range | Accuracy | Resolution | System Response Time | |
| Dissolved Oxygen | Membrane covered steady state polarographic | 0 to 500% air saturation 0 to 60 mg/L |
±2% or ±0.2 mg/L, whichever is greater (0–200%) ±6% of reading (200–500%) |
0.1% or 0.01 mg/L (0–10 mg/L) 1% or 0.1 mg/L (10–60 mg/L) |
98% of reading in 5 minutes |
| Temperature |
10,000 ohm @ 25 ºC thermistor | -0 to 45°C (32 to 113°F) |
±0.2°C | 0.1°C | |
| Conductivity | 4 electrode, Cell constant of 5.0 ±5% | 0-200 µS 0-2000 µs 0-20 mS 0-200 mS |
±0.5% of range (0-100 mS) ±1.0% of reading (100-200 mS) | 1.0 µS, 0.1mS Range dependent |
|
| Salinity | Calculated from conductivity and temperature | 0 to 80 ppt | ± 2 % or ± 0. 1 ppt | 0.1 ppt | |
| pH |
0 to 14 units | ±0.2 units | 0.01 units | ||
| ORP |
-1000 to +1000 mV | ±10 mV | 1 mV |
5200 Recirculating System Monitor Operations Manual
YSI Model 5567 DO Wiper/Stirrer
YSI 5200 Continuous System Monitor Specifications
Aquaculture Applications, Capabilities & Solutions
Español - Piscicultura Aplicaciones, Competencia y Soluciones
Aquaculture Monitoring & Control
Pacific White Shrimp Production in a Super-Intensive Raceway System
Super-Intensive Shrimp System Provides A Look Into The Future
High-Intensity Raceway Production Could Put Catfish Producers in the Fast Lane
Automation Streamlines High-Tech Shrimp Facility
Español - Automatización de alta tecnología lleva a la camaronicultura al siguente nivel
Renowned Newport Aquarium Uses YSI Instruments to Maintain Water Quality
Monitoring Recirculation Systems at Harbor Branch Oceanographic Institute with the YSI 5200
|
|
3682 Zobell ORP Calibration SolutionYSI 3682 Zobell solution is used for calibrating ORP/Redox on several YSI instruments. |
|
|
|
5580 Confidence Solution®Never worry about the accuracy of your pH, ORP, or conductivity data again! Ensure probe performance with the 5580 Confidence Solution®. |
|
|
|
5909 DO Cap Membrane KitYSI 5909 screw-on cap dissolved oxygen membranes. |
|