52 / 58
Portable Dissolved Oxygen Instruments
The portable YSI 52 is easy to use for field and lab work. An easy-to-read LCD provides clear, simultaneous display of time, temperature, % saturation and mg/L oxygen. Six D-cell batteries provide 100 hours of operation.
YSI has Authorized Distributors all over the globe. View a map and find your closest contact for YSI 52 / 58 products and accessories.
Have a question about your YSI 52 / 58? Check out the support page with resources compiled just for you.
Want to return a product for service? Here you’ll find authorized service centers and instructions on how to send in your YSI 52 / 58.
Use the Model 52 in the field with the field-proven 5739 dissolved oxygen sensor and connect a 5905 BOD probe in the lab to complete all of your lab testing. It's as simple as changing cables!
The YSI 58 is a proven field and laboratory instrument. The 58 displays oxygen concentrations in two temperature-compensated modes. The instrument offers manual salinity correction, recorder output, and operation from line power or batteries. The case also holds additional batteries for an optional DO field stirrer.
Please visit our extensive dissolved oxygen resource page to learn about all things DO and BOD related; www.YSI.com/weknowDO
YSI 52 Instrument General Specifications
| Dimensions | Size | 8.5 by 11 by 3.75 in (21.6 by 28 by 9.5 cm) |
| Weight | 4.3 lbs (2.4 kg) | |
| Operating Environment |
0 to 45°C, 10 to 90% humidity, non-condensing |
|
| Power | 6 alkaline D-cells batteries, or an AC adapter (optional). Batteries power the instrument for approximately 100 hours. |
|
| Salinity Compensation | Range: 0.0 to 40.0 ppt Accuracy: ±.02 mg/L |
|
| Temperature Compensation |
The mg/L mode is automatically temperature-compensated to an accuracy of ±1% of DO readings between 0 and 5°C, and to an accuracy of ±0.6% of readings between 5 and 45°C. The % air saturation mode is automatically temperature-compensated to an accuracy of ±0.5% of calibration values between 0 and 5°C, and to an accuracy of ±0.3% of values between 5 and 45°C. |
|
| Water Resistance |
With the probe receptacle capped, every case opening is has a gasket to resist the entry of water. |
|
| Warranty |
2-year instrument |
|
YSI 52 Specifications (Instrument Only)
| Range | Accuracy | Resolution | |
| Dissolved Oxygen (%) |
0 to 199.9% air saturation | ±0.1% of air saturation, plus 1 Least Significant Digit for % readings, plus probe error |
0.1% air saturation |
| Dissolved Oxygen (mg/L) |
0 to 19.9 mg/L | ±0.1% of saturation value, plus 1 Least Significant Digit for mg/L readings, plus probe error |
0.1% of the saturation value, expressed in mg/L, or 0.01 mg/L , whichever is greater |
| Temperature |
-5 to 45°C | ±0.1°C, plus probe error | 0.1°C |
YSI 58 Instrument General Specifications
| Medium | fresh, sea, or polluted water and most other liquid solutions | |
| Mode to Mode Accuracy (mg/L to % air saturation) |
+ 0.5% of reading, + 2 least significant digits (in the 0.01 mg/L mode) |
|
| Power | 4 D-size Alkaline Batteries (included) Approximately 1000 hours operation from each new set of batteries. Low battery indicator signal (LOBAT) appears automatically when approximately 50 hours of battery life remain. |
|
| Recorder Output | Voltage | 0 to 1 volt, full scale |
| Accuracy | + 0.4 % of full scale, + 1 least significant digit (in the 0.01 mg/L mode) |
|
| Minimum Load Impedance | 50 K OHM | |
| Salinity Compensation | Range | 0.0 to 40.0 ppt |
| Accuracy | + 0.3% of reading + 1 digit | |
| Temperature | Operating |
0 to +45 °C |
| Storage | 0 to +45 °C |
|
| Temperature Compensation |
The mg/L modes are automatically temperature compensated to an accuracy of + 1% of the dissolved oxygen reading between 5 and 45°C; and to an accuracy of +2% between 0 and 5. °C. See Temperature Sensitivity in probe specifications. |
|
| Water Resistance |
With RECORDER OUT, BATT ELIM and STIRRER receptacles capped, every case opening is splash resistant. | |
| Warranty |
2-year instrument |
|
YSI 58 Specifications (Instrument Only)
| Range | Accuracy | Resolution | |
| Dissolved Oxygen (%) |
0 to 200% air saturation | +0.3% Air Saturation | 0.1% air saturation |
| Dissolved Oxygen (mg/L) |
0 to 20 mg/L | 0.03 mg/L |
.01 mg/L |
| Temperature |
-5 to 45°C | ± 0.3 °C, plus probe interchangeability |
0.1°C |
Probe Specifications for:
YSI 5239
YSI 5718
YSI 5739
YSI 5750
YSI 5905
| Dissolved Oxygen Accuracy | ± 0.1 mg/L or ± 2% of reading, whichever is greater | |
| Electrolyte | 5739, 5718, 5750 |
KCl with Kodak Photo-Flo |
| 5905, 5239 | KCl with NA2SO4 | |
| Membrane Material | 5905 | FEP Teflon Cap |
| 5239, 5718, 5739, 5750 | .001” FEP Teflon Film .0005” FEP Teflon available, YSI 5776 |
|
| Pressure Compensation | Effective 0.5% of reading with pressures to 100 psi (230 feet of seawater) |
|
| Polarizing Voltage | 0.8 volts nominal | |
| Probe Current | Air at 30°C | 19 microamps nominal |
| Nitrogen at 30°C | 15 microamps or less |
|
| Probe Response Time | 5905 | Typical response to a step change in dissolved oxygen at 20°C is 90% of the change in 30 seconds. |
| 5239, 5718, 5739, 5750 |
Typical response for temperature and dissolved oxygen readings is 90% in 10 seconds at a low constant temperature of 30°C with the 5775 membrane. DO response and low temperature and low DO is typically 90% in 30 seconds. YSI 5776 High Sensitivity membranes may be used to improve response at low temperature and DO concentrations. |
|
| Sensor Materials | Cathode | Gold |
| Anode | Silver | |
| Temperature | Accuracy | + 0.2°C |
| Operating Range | 0° to 45°C | |
| Sensitivity | When measuring oxygen, the probe output current increases approximately 3.5% per 1°C of increase in temperature. The circuitry automatically compensates for this effect in a typical probe. However, the exact temperature sensitivity of an individual probe may vary slightly according to its condition. Therefore, a probe should be calibrated at a temperature close to the measurement temperature in order to minimize the possible effect of such variation. |
|
| Warranty |
1-year | |
58 Dissolved Oxygen Meter Operations Manual
5905, 5010, 200-BOD, and Pro-BOD Probe Instruction Manual
5239 Dissolved Oxygen Probe Instruction Manual
52 Dissolved Oxygen Meter Operations Manual
YSI 5905-W and 5010-W Wine Bottle Oxygen Probes
Water Quality Sampling for Field and Laboratory
How to Ensure Accurate and Successful BOD Measurements